Spectrophotometry Calibration Curves

INTOCHEMISTRY / INSTRUMENTAL ANALYSIS

Spectrophotometry and Calibration Curves

Use the slope and intercept of a linear calibration to calculate sample concentration and amount.

Learning objective

Rearrange A = mc + b, account for a fitted nonzero intercept, and assess when a concentration is within the calibrated range.

1. What the instrument measures

A spectrophotometer measures absorbance. In an appropriate linear working range at a fixed wavelength, a fitted calibration can connect absorbance A with analyte concentration c: A = mc + b. The fitted intercept b need not be zero.

2. Solve for concentration

Subtract the fitted intercept from sample absorbance, then divide by the slope: c = (A − b)/m. Check the units on the slope. A sample beyond the validated calibration range should be diluted and remeasured rather than extrapolated blindly.

Spectrophotometry and Calibration Curves concept and calculation diagram
Original, editable teaching diagram. Numerical values illustrate a worked case.

3. Worked example

For A = 0.240c + 0.035 with c in mM, a sample absorbance of 0.395 gives c = (0.395 − 0.035)/0.240 = 1.50 mM. In 15.0 mL of sample, amount = 1.50 mM × 15.0 mL = 22.5 µmol.

4. Quality checks

Use the same blank and wavelength conditions as the standards. Inspect whether standard points are reasonably linear and whether the sample falls within the range. Replicates, uncertainty, appropriate blanking and matrix effects matter for actual laboratory results.

Quick concept check

For A = 0.200c + 0.020, and A = 0.320, find c in mM.

Show worked answer

c = (0.320 − 0.020)/0.200 = 1.50 mM.